===== Telescope: ===== | ===== Telescope: ===== | ||
+ | ==== Focuser of the CDK20 does not run the complete range of ~35000μ ==== | ||
+ | |||
+ | <WRAP group> | ||
+ | <WRAP twothirds column> | ||
+ | **Problem description: | ||
+ | PWI3 shows clearly negative position values when driving in or the focuser does not run in the positive range (when driving out) up to the maximum ~35000μ. | ||
+ | |||
+ | **Solution: | ||
+ | The focuser has " | ||
+ | </ | ||
+ | <WRAP third column> | ||
+ | {{ ost: | ||
+ | </ | ||
+ | </ | ||
==== Wobbling Camera/ | ==== Wobbling Camera/ | ||
Line 210: | Line 224: | ||
===== Cameras: ===== | ===== Cameras: ===== | ||
+ | <WRAP group> | ||
+ | <WRAP twothirds column> | ||
==== Ice on the camera ==== | ==== Ice on the camera ==== | ||
- | |||
- | [{{ ost: | ||
**Description of the problem:** | **Description of the problem:** | ||
Line 218: | Line 232: | ||
**Solution: | **Solution: | ||
- | The cameras' | + | The cameras' |
==== ST-8: shutter error ==== | ==== ST-8: shutter error ==== | ||
**Description of the problem:** | **Description of the problem:** | ||
- | '' | + | //CCDops// reports '' |
**Solution: | **Solution: | ||
- | In this case the communication between the camera and CCDops is disturbed or the sensors for the shutter position do not work properly. Therefore, the software does not know whether the shutter is closed or not. It also could be that the shutter is indeed not moving as expected. In each case, it is often sufficient to switch off the camera and to disconnect it from the power supply. However, sometimes it is necessary to detach the camera from the telescope and to very carefully shake the camera a bit. Afterwards, the sensors of the shutter usually deliver proper data again. | + | In this case the communication between the camera and //CCDops// is disturbed or the sensors for the shutter position do not work properly. Therefore, the software does not know whether the shutter is closed or not. It also could be that the shutter is indeed not moving as expected. In each case, it is often sufficient to switch off the camera and to disconnect it from the power supply. However, sometimes it is necessary to detach the camera from the telescope and to very carefully shake the camera a bit. Afterwards, the sensors of the shutter usually deliver proper data again. |
+ | |||
+ | ==== Maxim DL: Two SBIG cameras cannot be simultaneously connected ==== | ||
+ | |||
+ | **Description of the problem: | ||
+ | When trying to couple two SBIG cameras simultaneously with //Maxim DL//, //Maxim DL// becomes very slow, hangs or crashes. | ||
+ | |||
+ | **Solution: | ||
+ | This can be due to the fact that '' | ||
+ | |||
+ | </ | ||
+ | <WRAP third column> | ||
+ | [{{ ost: | ||
+ | \\ | ||
+ | \\ | ||
+ | \\ | ||
+ | [{{ ost: | ||
+ | </ | ||
+ | </ | ||
==== CCDops does not find the filter wheel ==== | ==== CCDops does not find the filter wheel ==== | ||
**Description of the problem:** | **Description of the problem:** | ||
- | '' | + | //CCDops// does not find the filter wheel of the STF8300 camera although the filter wheel is rotating when the power is switched on. |
**Solution: | **Solution: | ||
- | The filter wheel needs to be reinitialized. The corresponding menu can be found via //Filter// and //Filter Setup//. Select the " | + | The filter wheel needs to be reinitialized. The corresponding menu can be found via '' |
Line 244: | Line 276: | ||
- | ====== | + | ====== |
++++ GM 4000 Mount: Tracking error | | ++++ GM 4000 Mount: Tracking error | |