Description of the problem: The current alignment has been lost, e.g. by an incorrect operation (delete Alignment).
Solution: Restore a previously saved alignment. If such an alignment exists, it can be selected with:
MENU - Alignment - Align Database - Load Model
If there is no saved alignment, you will have to create a new alignment.
Description of the problem: The telescope does not point to the selected objects or to wrong ones.
Solution: Possible solutions are in this order:
Description of the problem: The tracking speed does not automatically adjust to the target, i.e. the moon.
Solution: By default, the following option should be enabled:
Menu - Drive - Follow Obj
If it is not, the automatic switching between different tracking modes does not work.
Description of the problem: The telescopes does not move to the default park position (pointing to the northern celestial pole).
Solution: The solution is to switch back to the default park position via:
Menu - Alignment - Park Position - Default park
Description of the problem: The dome's wall obscures the view of the telescope to the sky. This happens when the dome does not follow the pointing of the telescope, i.e. the open shutter is not in phase with the telescope.
Solution: The dome control can determine the current position of the dome by means of a magnet. For this purpose, the dome must be moved through the south point once:
Dome Tracks Scope
.Postions
section. For the latter, the desired target azimuth (Azimuth
) must be selected and then Go To
clicked. Description of the problem: The motor that drives the dome rotation can be heard but the dome does not rotate.
Solution:
Dome Tracks Scope
.Description of the problem: The shutter of the dome does not close completely and a small gap remains between the shutter and the hatch.
Solution: Open the shutter until it is possible to move the hatch. Open the hatch a bit and close it again. Afterwards, it should be possible to completely close the shutter.
Description of the problem: The shutter or the hatch do not close via the remote control.
Solution 1: On the hand terminal, use the command to close the dome completely:
MENU - Settings - Dome - Close Shutter
. After that, the hatch and shutter should open and close properly again using the remote control.
Solution 2:
In Maxim DL in the Observatory control window go to the Dome-Tab and try to open the dome again with Open
. Afterwards close it with Close
.
Problem description: The Sshutter and/or hatch do not move smoothly when controlled by the remote control.
Solution: This is normal when the dome is coupled to Maxim DL at the same time. In this case the signals coming from the remote control overlap with the signals coming from Maxim DL. If the dome is to be controlled manually, it is recommended to disconnect it from Maxim DL. However, if there is currently no connection to Maxim DL, there might be a serious problem. In this case, immediately contact the technical staff.
Problem description: The shutter and/or hatch can no longer be moved even with the methods described above, because a component may have broken.
Solution:
Problem description: PWI3 shows clearly negative position values when driving in or the focuser does not run in the positive range (when driving out) up to the maximum ~35000μ.
Solution:
The focuser has “swallowed” and the encoder delivers wrong values. This can be fixed by clicking on Home
(see picture) in the Focuser tab of PWI3 under Find Home and Reset
and then on YES
. This will reset the focuser and reinitialize the home position, restoring the normal focus range.
Description of the problem: You may find that the camera or spectrograph still wobble or rotate despite your efforts to secure them to the telescope.
Solution: The screws for attaching devises could be rusty and therefore, it is hard to tighten them. Thus, the screws should be completely unscrewed. After checking them they should be cleaned or if they are too damage, they should be exchanged. In general it can't be bad to unscrew them from time to time to check them. In the worst case the screw thread is damaged and then it has to be cleaned or the thread needs a new cut.
Description of the problem: The hand terminal shows the error message No Response 16 or No Response 17.
Solution: These error codes appear in case the communication between control unit and engines of the right ascension axis and/or declination axis is disturbed. First, check if all cables are correctly plugged in. If that didn't solve the problem, use contact spray on the connectors of the plugs (attention: Do not directly spray into the sockets). If the problem persists, the hand terminal is probably broken.
Description of the problem: The sockets for the right ascension and the declination cable of the electronic box are loose and it is not possible to tighten them.
Solution: The bolts that hold the sockets are countered from inside the box by means of nuts. Therefore, it is inevitable to open the box. The cover of the box is bolted into the upper part of it. Since the cover is plain with the exception of a few boreholes, it could be a bit troublesome to remove it. One can try to turn the cover by means of a screwdriver but one has to be very careful in order not to damage the electronics. However, the best option is to build a special tool for this purpose. In each case a certain amount of strain is necessary to open the electronic box. Therefore, it probably is a good idea to get some help, so that somebody or something can hold the box while turning the cover. When the box is open, the bolts of the sockets can be easily tightened. Subsequently, the electronic box can be closed again. It is important to carefully tighten the cover. Otherwise, it might be not possible to attach the mechanics to the electronic box when the mount is set up.
Description of the problem: Instead of driving towards objects correctly and precise, the telescope points to the ground. Furthermore, the movements of the telescope (along larger distances) are (very) slow.
Solution: It's likely that the power supply is causing the problem. Try a different power supply.
Description of the problem: The CCD chip of the camera has experienced icing, see image on the right.
Solution: The cameras' interior is not in a vacuum and moisture can get in, causing icing on the cooled CCD chip. To prevent this, the cameras have dry cartouches with desiccants. If icing appears, the cartouches need to be regenerated (see this article). During that observational night, wait for approximately one hour until the ice sublimates from the CCD chip.
Description of the problem:
CCDops reports shutter error
while taking the first images after switching on the camera.
Solution: In this case the communication between the camera and CCDops is disturbed or the sensors for the shutter position do not work properly. Therefore, the software does not know whether the shutter is closed or not. It also could be that the shutter is indeed not moving as expected. In each case, it is often sufficient to switch off the camera and to disconnect it from the power supply. However, sometimes it is necessary to detach the camera from the telescope and to very carefully shake the camera a bit. Afterwards, the sensors of the shutter usually deliver proper data again.
Description of the problem: When trying to couple two SBIG cameras simultaneously with Maxim DL, Maxim DL becomes very slow, hangs or crashes.
Solution:
This can be due to the fact that Camera threading on (recommended)
is set in the camera options. This option must be deactivated for the SBIG cameras. Afterwards the cameras should connect and work as usual.
Description of the problem: CCDops does not find the filter wheel of the STF8300 camera although the filter wheel is rotating when the power is switched on.
Solution:
The filter wheel needs to be reinitialized. The corresponding menu can be found via Filter
and Filter Setup
. Select the “FW-5” filter wheel from the drop down menu and confirm the settings with a click on the OK
button. Afterwards the filter wheel should work again.
Current user's manuals are in the laboratory course room (2.009):
Manual | Comment |
---|---|
10Micron GM 4000 QCI mount | Blue folder |
10Micron GM 2000 QCI mount | Folder (mount not available) |
Celestial Globe | Folder |
VNC | Folder |
T-Point | Software hand book |
PC-Control of the mounting GM 2000 | Folder, for “Perseus”, “The Sky”, and “Cartes du Ciel” |
SBIG Users Guide: CCDOps | Ring bond, for CCDOps Version 5 |
SBIG Operating Manual CCD Cameras | Ring bond, for ST-7XE, ST-8XE, ST-9XE, ST-10XE, ST-10XME, and ST-2000XM |
SBIG CCD-Camera manual (german) | Folder, Baader, for cameras and control/data reduction software |
SBIG AO-7 Manual (1997) | Loose paper collection, Manual for adaptive optics |
SBIG AO-7 Manual (2003) | Loose paper collection, Manual for adaptive optics |
Manual for automatic observation domes | Folder, from Baader |
TCF-S Manual | Folder, (printed from CD?) manual for focusing |
Visual Spec Manual | Folder, data reduction software for spectra (Windows) |
DADOS Manual | Folder for the spectrograph |
Celestron Advanced Series GT Manual | Manual for the mobile telescope |
Folder with various short manuals | e.g. Micro-Guiding-Crosshair-Ocular, USB-equipment |
Panasonic Toughbook CF-31 | Manual |